The
performance of OraQuick, a widely used rapid point-of-care test for HIV
diagnosis, is slightly poorer when testing oral fluid samples than when testing
blood samples, according to a meta-analysis published in the online edition of
Lancet Infectious Diseases this week.The authors also found that if the test
is used in low-prevalence settings with oral fluid samples, over one-in-ten reactive results may turn out to be false positives.
Rapid diagnostic
tests, which can be operated at the point of care by doctors or nurses, have a
number of attractive features in comparison with conventional laboratory tests.
Such devices are essential in resource-limited settings where laboratories are
inaccessible or unavailable. As patients do not need to come back for their
results on another day, far more people receive their results.
Rapid tests
also have the advantage of not requiring a blood sample taken with a
needle. Most require a tiny sample of whole blood, taken by pricking a
fingertip with the sharp point of a lancet. Some tests can be used with oral
fluid, obtained by swabbing an absorbent pad around the outer gums, adjacent to
the teeth.
The OraQuick
Advance Rapid HIV-1/2 is, by far, the most widely used test which can test oral
fluid samples. It is also one of the rapid tests which has been most
extensively evaluated by independent researchers. Whilst accuracy has generally
been reported to be very good, a few reports have identified problems including false positive results and
limited sensitivity in people with recent infection. Such problems seem to have
been associated in part with the use of oral fluid samples.
Dr Nitika
Pant Pai of McGill University and colleagues conducted a systematic review and
meta-analysis in order to identify and pull together all the studies that have
been conducted on the use of OraQuick in clinical practice.
Twenty-four
publications were identified, approximately half from the United States and
half from African or Asian settings. Some of these studies were very
large, and just under 40,000 samples were tested in total.
The key measures of accuracy in such studies are sensitivity (the percentage of results
that are correctly positive when HIV is actually present) and specificity (the percentage of results
that are correctly negative when HIV is not present).
The meta-analysis confirmed the very high sensitivity and
specificity of OraQuick when testing blood samples.
The primary
objective of the meta-analysis was to compare the test’s performance with oral
fluid and blood samples. In seven studies which made a direct comparison
of performance on these two types of sample, the pooled sensitivity with oral
fluid was lower than with blood – a difference of around 2%. In contrast, the
difference in pooled specificity was not so marked.
While the
differences may be relatively small and potentially compensated by the high
acceptability of testing with oral fluid, it is important to note the 95%
confidence intervals when reading the following figures. Confidence intervals
represent plausible estimates within which the ‘true’ figure may lie – for
example, sensitivity could be as low as the lower figure given.
| |
Sensitivity (95%
confidence interval)
|
Specificity (95%
confidence interval)
|
|
Oral fluid
|
98.03% (95.85 –
99.08%)
|
99.74% (99.47 -
99.88%)
|
|
Blood
|
99.68% (97.31 -
99.96%)
|
99.91% (99.84 -
99.95%)
|
The
researchers also pooled data from six studies which tested oral fluid only and
eleven studies which tested one form or another of blood sample. As these involve
different study populations, the comparison is somewhat less reliable, but
similar results were found.
The differences in performance will have a particular impact when the test is used in populations in which HIV prevalence is relatively low. As with other tests, a few people who test will have a false positive result. In a setting where very few people have HIV, the majority of apparent positive results will in fact be incorrect. However, as the proportion of people with HIV being tested increases, the true positives start to outnumber false positives.
The authors found that when testing oral fluid in a population where prevalence is below 1%, the positive predictive value (the proportion of individuals with positive test results who are correctly diagnosed) was 88.6% (95% confidence interval 77.3 - 95.9%). On the other hand, in populations with HIV prevalence above 1%, positive predictive value with oral fluid was 98.6% (95% confidence interval 85.7 - 99.9%) and with blood 98.5% (93.1 - 99.8%).
These data
derive from evaluations of OraQuick as used in clinical practice and are lower
than the manufacturer’s claims. However almost all data from manufacturers are
derived from assessments in carefully controlled laboratory settings.
The authors
examined reports of false positive results in order to identify factors that
could have caused them: errors in test performance and conduct of test (i.e.
inaccurate specimen collection, gum swabbing more than once), and
errors in the interpretation of results (interpreting weakly reactive lines).
These were generally linked to inadequate training of staff or lapses in
quality assurance.
However
Nitika Pant Pai says that the key reason for poorer performance with
oral fluid is that there are lower quantities of HIV antibodies in oral fluid than in whole
blood. The quantities of antibodies in oral fluid are especially low after
recent infection.
The authors
suggest that people who think they may have recently been exposed to HIV should
treat negative oral fluid results with caution. Similarly, people who are at
higher risk of HIV infection who get a negative result should seek further
testing.
On the
other hand, if the test is used in low-prevalence settings, there will be
a higher number of false positive results than with blood-based testing. As
with any HIV test, an apparently positive result must be confirmed with
supplementary testing.